How do I use the auto calibration feature in DinoCapture?

Auto Calibration requires a Dino-Lite Edge Series microscope camera, CS-40 or CS-41 Auto Calibration slide, and DinoCapture 2.0 versions 1.5.25 or newer, click here to download the latest version.

Auto Calibration provides an easier way to calibrate a Dino-Lite Edge Series microscope for measurement. The DinoCapture 2.0 software identifies special QR codes on the CS-40 / CS-41 slide and performs the calibration automatically.

For older calibration slides (CS-30, MS5MM) or seeking manual calibration information:  PDF guide | Video guide


CS-41 Auto-Calibration Target, enlarged to show details.
CS-40 Auto-Calibration Target, enlarged to show details.

QR codes are divided into 3 rows:

  • Row A: magnifications of 120x and above
  • Row B: magnifications of 40x – 220x
  • Row C: magnifications of 20x – 50x

Tip: For more accurate results, it is recommended to create and save a calibration profile for each magnification range that measurements will be performed at. Calibration profiles are stored in the Calibration drop-down menu and a check mark will appear next to the selected profile.

Automatic calibration with the CS-40 / CS-41 slide

Instruction Screenshot
Bring the CS-40 / CS-41 into focus with the Dino-Lite. Tips: Make sure the microscope is perpendicular to the calibration target. View two or more QR codes for the best results. Misaligned or heavily tilted slide while in view may result in calibration failure Screenshot
Click the Calibration tool menu item and select New Calibration Profile. Type in a name for this calibration profile and click Continue Calibration. Screenshot
Press F8 or click on the Freeze button. Screenshot
Input the Magnification as shown on the dial. Microscopes with AMR display the magnification automatically. Press Enter. Screenshot
The software will automatically perform the calibration measurement and provide the adjusted magnification. Click Finish to complete the calibration. Screenshot
Verify your calibration by performing a test measurement with a known standard or the 10mm scale on the CS-40 / CS-41 slide Screenshot
Tips: For best results, select a calibration profile close to the magnification you are using. If fixed magnification is required, simply calibrate at that magnification only for the most accurate results. Calibration can be disabled by selecting No Calibration in the drop-down menu. Screenshot

need more help?

Dino-Lite Europe, IDCP B.V.
Manuscriptstraat 12-14
1321 NN Almere
The Netherlands
+31(0)20-6186322

Sign up for
Dino-Lite newsletter


The official main distributor and importer for Europe, the Middle-East and Africa. Privacy statement and General Conditions.  To withdraw consent use this form copyright © IDCP B.V. - The Netherlands

Dino-Lite Europe, IDCP B.V.
Manuscriptstraat 12-14
1321 NN Almere
The Netherlands
+31(0)20-6186322

Sign up for
Dino-Lite newsletter


The official main distributor and importer for Europe, the Middle-East and Africa. Privacy statement and General Conditions.  To withdraw consent use this form copyright © IDCP B.V. - The Netherlands

Dino-Lite Europe
IDCP B.V.
Manuscriptstraat 12-14
1321 NN Almere
The Netherlands
+31(0)20-6186322

Sign up for
Dino-Lite newsletter



The official main distributor and importer for
Europe, the Middle-East and Africa. 
Privacy statement and General Conditions. 

To withdraw consent use this form
copyright © IDCP B.V. - The Netherlands